Surgical Sponge Count
Simulation with photos from the test set: table with 14 sponges and then with 13 · Counts the surgical sponges on the table or tray and warns when the count does not match the one opened at the start
- No alerts so far
About this solution
For operating rooms and hospitals. The camera above the back table or the tray of used sponges counts the sponges all the time and warns when the count at the end does not match the one opened at the start, as support for the nursing count against a sponge left in the patient.
- 1Use a camera above the back table or the tray where used sponges are laid out, and mark that area. The camera does not need to see the patient.
- 2The model, trained by Neural with photos of clean and bloody sponges on surgical drapes, finds each sponge in the image.
- 3The count is the median of the last few seconds: a hand or instrument passing over does not change the number. Each new steady count is recorded with the time.
- 4Enter how many sponges were opened: if the steady count differs from that, the warning "Count different from expected" comes up, to recount before closing.
- 5It supports the team's count and does not replace the safe surgery protocol check. Sponges piled on top of each other are not counted one by one: they need to be laid out separately on the tray.
- 6Tested on photos the model did not see in training, from above the table: it found 99.4% of the sponges. In videos shot from the side with a moving camera the count is not usable, and forceps shining under the surgical light even looked like sponges: that is why the camera stays fixed, from above.
Training data: Roboflow Universe — gauze (GAUZE IMAGES), gauze (GAUZE IMAGES, grade de contagem), New Workspace (Gauze) (CC BY 4.0).
Benefits
- Warning to recount before closing when the count at the end differs from the one opened at the start.
- Record of each count with the time, for the safe surgery checklist audit.
- Uses one camera above the back table; it does not film the patient. Support for nursing, not a replacement for the count.
More demos

Checks that handwashing happens and lasts the minimum time

Tells standing, sitting and lying on the floor apart to flag falls without false alarms

Tracks patient position and alerts when it is time to reposition

Occupational safety regulations — healthcare workers without rings, watches and bracelets

Gown/coverall, mask, gloves, goggles and face shield in patient care areas